Description
General Specification
Brand | Microspec |
Model | Titra+ |
Material | Mild Steel |
Automatic Grade | Automatic |
Voltage | 220-240 V |
Frequency | 50-60 Hz |
Features | Accurate analysis of silver using chemicals – Automatically |
I deal in | New Only |
Electrical Data
Power supply and consumption | AC 115 or 230 V, 50/60 Hz, max. 120 W without evaluation PC |
Protection class | IP40 |
Environmental Conditions
Operating temperature | 10 °C – 45 °C / 50 °F – 113 °F |
Storage/Transport temperature | -20 °C – 80 °C / -4 °F – 176 °F |
Relative humidity | ≤ 95 % |
Sample Alignment
Sample positioning | Manually |
Video microscope | High-resolution CCD colour camera for optical monitoring of the measurement location along the primary beam axis, Crosshairs with a calibrated scale (ruler) and spot-indicator, Adjustable LED illumination |
Zoom factor | Digital 1x, 2x, 3x, 4x |
Evaluation Unit
Computer | Windows®-PC |
Software | WinFTM® optimized for GOLDSCOPE Including Gold Setup, Jewelry Setup and Fashion Jewelry Setup |
Standards
CE approval | EN 61010 |
X-Ray standards | DIN ISO 3497 and ASTM B 568 |
Approval | Individual acceptance inspection as a fully protected instrument according to the German regulations “Deutsche Röntgenverordnung – RöV”. |
Sample Stage
Design | Fixed sample support |
Max. sample weight [kg/lb] | 13/29 |
Usable sample placement area [mm/in] | 305 x 490/ 12 x 19.3 |
Max. sample height [mm/in] | 130/5.1 |
Dimensions
External dimensions | 405 x 588 x 426/ |
Width x depth x height [mm/in] | 16 x 23 x 17 |
Weight [kg/lb] | approx. 45/99 |
X-Ray Source
X-ray tube | Tungsten tube, thermally stabilized |
High voltage, three steps [kV] | 30, 40, 50 |
Primary filter, Material and thickness [μm/mils] | none |
Aperture (Collimator) | fixed |
Ø [mm/mils] | 1.0/39 |
X-Ray Detection
Detector type | Silicon PIN detector peltier-cooled |
Resolution fwhm for Mn-Kα [eV] | ≤ 180 |
Element range | S (16) to U (92) |
Measuring distance [mm/in] | 0 – 25/0 – 1, Distance compensation with patented DCM method for simplified measurements at varying distances. For particular applications or for higher demands on accuracy an additional calibration might be necessary. |
Reviews